| SPECIFICATIONS |
| MODEL |
5130
|
5210
|
| AC WITHSTAND VOLTAGE |
| Power |
300W
|
150W X 2
|
| Current |
60A
|
30A
|
| Voltage |
60V
|
60V
|
| Min.Operation Voltage(DC) |
1.0V at 60A
|
1.0V at 30A
|
| CONSTANT CURRENT MODE (CC MODE) |
| Range |
0-60A
|
0-6A
|
0-30A
|
0-3A
|
| Resolution,12bit |
15mA
|
1.5mA
|
7.5mA
|
0.75mA
|
| Accuracy |
±(0.1% of setting + 0.1% F.S.)
|
±(0.1% of setting + 0.1% F.S.)
Uunder 100W
|
±(0.1% of setting + 0.1% F.S.)
|
| CONSTANT VOLTAGE MODE (CV MODE) |
| Range |
1-60V
|
1-60V
|
| Resulotion,12bit |
15mV
|
15mV
|
| Accuracy |
±(0.05% of setting + 0.1% F.S.)
|
±(0.05% of setting + 0.1% F.S.)
|
| CONSTANT RESISTANCE MODE (CR MODE) |
| Range |
1Ω~4KΩ (300W/60V)
|
25mΩ~1Ω (300W/60V)
|
2Ω~8KΩ (300W/60V)
|
50mΩ~2Ω (300W/60V)
|
| Resolution,12bit |
25m(1/Ω)
|
25mΩ
|
0.125m(1/Ω)
|
0.5mΩ
|
| Accuracy |
±(0.2% of setting + 0.2% F.S.)
|
±(0.2% of setting + 0.2% F.S.)
|
| DYNAMIC MODE (Transient Operation) |
| Dynamic Mode |
CC MODE
|
CC MODE
|
| T1 & T2 or (THigh/TLow), 10KHZ |
1mS~20S
|
0.025mS~10mS
|
1mS~20S
|
0.025mS~10mS
|
| Resolution |
1mS
|
1uS
|
1mS
|
1uS
|
| Slew Rate |
0.01A~2.5A/uS
|
1mA~250mA/uS
|
0.005A~1.25A/uS
|
0.5mA~125mA/uS
|
| Resolution |
10mA/uS
|
1mA/uS
|
5mA/uS
|
0.5mA/uS
|
| Current |
0-60A
|
0-6A
|
0-30A
|
0-3A
|
| Resolution |
15mA
|
1.5mA
|
7.5mA
|
0.75mA
|
| Accuracy |
±(0.3% of setting + 0.3% F.S.)
|
±(0.3% of setting + 0.3% F.S.)
Under 100W
|
±(0.3% of setting + 0.3% F.S.)
|
| MEASUREMENT SECTION |
| Voltage Read Back |
| Range |
0-60V
|
0-60V
|
| Resolution,12bit |
1mV
|
1mV
|
| Accuracy |
±(0.02% of reading + 0.02% F.S)
|
±(0.02% of reading + 0.02% F.S)
|
| CURRENT READ BACK |
| Range |
0-60A
|
0-6A
|
0-30A
|
0-3A
|
| Resolution,12bit |
1mA
|
0.1mA
|
0.5mA
|
0.05mA
|
| Accuracy |
±(0.1% of reading + 0.1% F.S.)
|
±(0.1% of reading + 0.1% F.S.)
Under 100W
|
±(0.1% of reading + 0.1% F.S.)
|
| PROTECTION |
| Current Limit (On CV &CR Mode) |
| Range |
0-60A
|
0-30A
|
| Resolution ,12bit |
15mA
|
7.5mA
|
| Accuracy |
±(1% of setting + 1% F.S.)
|
±(1% of setting + 1% F.S.)
|
| General Protect |
| Over Power Protection |
≒315W
|
≒157.5W
|
| Over Current Protection |
≒63A
|
≒31.5A
|
| Over Voltage Protection |
≒63V
|
≒63V
|
| Over Temperature Protection |
≒85℃
|
≒85℃
|
| Reverse Polarity |
60A (during 1min)
|
30A (during 1min)
|
| GENERAL |
| Voltage Trigger ON/OFF |
| Range |
0-60V
|
0-60V
|
| Resolution ,12bit |
15mV
|
15mV
|
| Accuracy |
±(0.5% of reading + 0.5% F.S)
|
±(0.5% of reading + 0.5% F.S)
|
| Short Circuit MAX Current |
≒60A
|
≒30A
|
| Short Circuit Resistance |
≒15mΩ
|
≒30mΩ
|
| I Monitor |
0 - 10V
|
0 - 10V
|
External Control
(Via mainframe model 5101 or 5102) |
0 - 10V
|
0 - 10V
|
| Dimension |
104 (W) x 143(H) x 470(D) mm
|
104 (W) x 143(H) x 470(D) mm
|
| Weight |
6.2Kg
|
6.3Kg
|